EasyManua.ls Logo

Sartorius Cubis MCE - Metrological Data; Models MCE225 S | MCE225 P | MCE125 S | MCE125 P

Sartorius Cubis MCE
98 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Operating Instructions Cubis
®
MCE 85
Technical Data
15.7 Metrological Data
15.7.1 Models MCE225S | MCE225P | MCE125S | MCE125P
MCE225S MCE225P MCE125S MCE125P
Unit Value Value Value Value
Scale interval (d) mg 0.01 0.01 | 0.02 | 0.05 0.01 0.01 | 0.1
Maximum capacity (Max) g 220 60 | 120 | 220 120 60 | 120
Repeatability at 5% load
Standard deviation of the load values,
tolerance
mg 0.015 0.015 0.015 0.015
Standard deviation of the load values,
typical value
mg 0.0065 0.0065 0.0065 0.0065
Repeatability at approx. maximum capacity
Standard deviation of the load values,
tolerance
mg 0.025 0.04 0.025 0.06
Standard deviation of the load values,
typical value
mg 0.02 0.02 0.02 0.02
Linearity deviation
Tolerance mg 0.1 0.15 0.1 0.15
Typical value mg 0.065 0.1 0.065 0.1
Deviation when load is off-center, positions
according to OIML R76
Test weight g 100 100 50 50
Tolerance mg 0.15 0.2 0.15 0.2
Typical value mg 0.1 0.1 0.1 0.1
Sensitivity drift between +10°C and +30°C ppm/K 1 1 1 1
Tare-maximum capacity: Less than 100% of
maximum capacity
Accuracy class according to Directive
2014/31/EU
I I I I
Verification scale interval (e) according to
Directive 2014/31/EU
mg 1 1 1 1
Minimum load (Min) according to Directive
2014/31/EU
mg 1 1 1 1
Minimum initial weighing according to USP
(United States Pharmacopeia), Chap. 41
Optimum minimum initial weighing mg 8.2 8.2 8.2 8.2
Typical minimum initial weighing mg 13 13 13 13
Typical stabilization time s 2 2 2 2
Typical measurement time s 6 6 6 6

Table of Contents