52 Operating Instructions Cubis
®
MCE
System Settings
Parameters in the “USB” Submenu
Parameters Setting values Explanation
DEV.USED NONE* Indicates that no device connection is detected at the USB port.
[Device
Designation]
Displays the designation of the devices connected to the USB port.
* Factory setting
7.3.3 Parameters in the “Data Output” Main Menu
Parameters in the “SBI Communication” Submenu
Parameters Setting values Explanations
COM. OUTP. IND.NO* Starts the data output after pressing the key or software command as an
individual value without stability.
IND.AFTER Starts the data output after pressing the key or software command as an
individual value after stability.
AUTO W/O Activates the automatic data output without stability.
AUTO W/ Activates the automatic data output after stability.
STOP.AUTO OFF* Deactivates the option to abort the automatic data output.
ON The automatic data output is aborted by pressing the [PRINT] key or a
software command.
AUTO.CYCL. EACH VAL* Starts the automatic data output with a cycle after each value.
2. VALUE Starts the automatic data output with a cycle after every 2nd value.
INTERV. Starts the automatic data output at the output rate set under “INPUT /
INTERV.”
FORMAT 16 CHARS Data output provides 16 characters per line (16 characters for the measured
values only).
22 CHARS Data output provides 22 characters per line (16 characters for the measured
values and 6 characters for identifiers).
EXTR.LINE Data output provides an additional line with the date, time, and weight
value.
AUTO.TARE OFF* Deactivates automatic taring after data output.
ON The device automatically tares after data output.
* Factory setting