Operating Instructions Cubis
®
MCE 53
System Settings
Parameters in the “Settings for Printouts” Submenu
Parameters Setting values Explanation
ACTIVATE MAN. NO Manual without stability: Print process can be started manually at any time.
MAN.AFTER* Manual after stability: After pressing the [Print] key, the print command is
only executed once stability is achieved.
INTERV. Starts the automatic data output at the output rate set under “INPUT /
INTERV.”
AUTO.LC Automatically at load change: Print process starts after every load change.
FORMAT 22 CHARS* Printer output prints 22 characters per line (16 characters for the measured
values and 6 characters for identifiers)
EXTR.LINE Printer output prints an additional line with the date, time, and weight value.
PRINT.INIT OFF Deactivates the output of the application parameters.
ALL* Print command prints all parameters.
MAIN PAR. Print command only prints the main parameters.
GLP OFF* Deactivates the GLP printout.
CAL.-ADJ. Activates the GLP printout for all calibration and adjustment processes.
ALWAYS − The GLP printout is always switched on.
− All printouts contain a GLP header and a GLP footer.
TAR./PRT. OFF* Deactivates automatic taring after printer output.
ON Automatically tares the device after every printout.
TIME 24H* Sets the time to 24-hour mode.
12H − Sets the time to 12-hour mode (AM | PM).
− Locked when date format YYYY.MM.DD (ISO) is activated.
DATE DD.MMM.YY* Sets the date display format to DD.MMM.YY
MMM.DD.YY Sets the date display format to MMM.DD.YY
YYYY.MM.DD Sets the date display format to YYYY.MM.DD (ISO).
* Factory setting
Parameters in the “PC Direct Transfer” Submenu
Parameters Setting values Explanations
DEC.SEPAR. DEC.POINT* Sets a point as a decimal separator.
DEC.COMMA Sets a comma as a decimal separator.
OUTP.FORM. TEXT+NUM.* Exports text and numbers.
ONLY NUM. Only exports numbers.
* Factory setting