62 Operating Instructions Cubis
®
MCE
Operation
8.8 Internally Calibrating and Adjusting the
Device
Calibration and Subsequent Adjustment
Requirements
− The weighing pan is unloaded.
− The operating display shows a stable weight value.
Procedure
t To set the internal adjustment in the menu: In the “SETUP” / “BALANCE”
menu, for the “CAL./ADJ.” parameter, select the “INT.CAL.”
configuration value.
t To set the sequence of the adjustment function to calibration and
subsequent automatic adjustment: In the “SETUP” / “BALANCE” menu,
for the “CAL.SEQ.” parameter, select the “ADJUST” configuration value.
t Exit the menu.
t Press the [Zero] key.
t To start the internal calibration | adjustment: Press the [Adjust] button.
t If automatic leveling is set for a model with motorized adjusting feet: The
device levels itself automatically.
y The internal calibration weight is loaded.
y The [
CAL.RUN.
] display appears in the operating display during
calibration.
y The device is automatically calibrated and the internal calibration weight
is removed.
y The [CAL.END] display indicates the end of calibration.
Manually Starting Adjustment after Calibration
Requirements
− The weighing pan is unloaded.
− The operating display shows a stable weight value.
Procedure
t To set the internal adjustment in the menu: In the “SETUP” / “BALANCE”
menu, for the “CAL./ADJ.” parameter, select the “INT.CAL.”
configuration value.
t To set the sequence of the adjustment function to calibration without
subsequent automatic adjustment: In the “SETUP” / “BALANCE” menu,
for the “CAL.SEQ.” parameter, select the “CAL/ADJUST” configuration
value.
t Exit the menu.
t Press the [Zero] key.
t To start the internal calibration: Press the [Adjust] button.
t If automatic leveling is set for a model with motorized adjusting feet: The
device levels itself automatically.
y The internal calibration weight is loaded.
y The
[
CAL.RUN.
]
display appears in the operating display during
calibration.
y The calibration weight value is shown.