Operating Instructions Cubis
®
MCE 67
Operation
t To switch the displayed mass unit during weighing or before starting an
application:
t If this relates to a device with a motorized draft shield: Press the
[Toggle between weight units] button multiple times if necessary,
until the desired unit is displayed.
t If this relates to a device without a motorized draft shield: Press the
[Toggle between weight units] button or the [Change] key multiple
times if necessary, until the desired unit is displayed.
y The current weight value is displayed in the selected unit.
8.13.2 Running the “Statistics” Application
The “Statistics” application saves up to 99 weight values and evaluates these
statistically.
The following values are saved and exported by the statistics application:
− Number of components
− Mean value
− Standard deviation
− Variation coefficient
− Sum of all values
− Lowest value (minimum)
− Highest value (maximum)
− Spread: Difference between maximum and minimum
The “Statistics” application can be combined with the following functions:
− Toggle between weight units only available before starting the applica-
tion
− ID marking
− Automatic taring
Procedure
t Select the Statistics application in the “APPLIC.” menu.
t Exit the settings menu.
t Place a sample on the weighing pan.
t Change the displayed weight unit if necessary.
t To start the recording of the statistics: Press the [Confirm] button on
the keypad.
y The current weight value is saved.
t To save the next value:
t Remove the sample on the weighing pan.
t Place a new sample on the weighing pan and press the [Confirm]
button.