66 Operating Instructions Cubis
®
MCE
Operation
8.12 Turning the Ionizer On | Off (Optional)
Procedure
t To activate the ionizer in the menu and to determine the intensity at
which the ionizer is to operate: In the “DEVICE” / “IONIZER” menu, for
the “POWER” parameter, select the “SOFT”, “MEDIUM” or “STRONG”
configuration value.
t To specify the duration of the ionization process in the menu: In the
“DEVICE” / “IONIZER” menu, for the “DURATION” parameter, select a
value of “3 SEC” to “15 SEC”.
t Exit the settings menu.
y The [Ionizer] button appears in the operating display.
t To start the ionization process: Press the [Ionizer] button.
t To switch off the ionizer in the menu: In the “DEVICE” / “IONIZER” menu,
for the “POWER” parameter, select the “OFF” configuration value.
8.13 Running Applications (Examples)
8.13.1 Executing the “Toggle Between Weight Units” Function
The “Toggle between weight units” function enables a switch between the
basic unit display and up to 4 other units. The units can be set during the
weighing process and the decimal places can be adjusted.
The basic unit is displayed every time the device is started.
Procedure
t To activate the “Toggle between weight units” function in the menu: In
the “BALANCE” menu, for the “UNIT” parameter, select the “ON” con-
figuration value.
t Exit the settings menu.
y The device boots up.
t To select the convertible units:
t If this relates to a device with a motorized draft shield: Press and hold
down the [Toggle between weight units] button.
t If this relates to a device without a motorized draft shield: Press and
hold down the [Toggle between weight units] button of the [Change]
key.
t The “Toggle between weight units” function menu opens.
t Select a submenu from “Unit 1” – “Unit 4”.
t Select the desired unit and confirm the display (available units, see
Chapter “7.3.7 Parameters in the “Unit 1”, “Unit 2”, “Unit 3”, and “Unit 4”
Application Menus”, page 57).
t To set the number of display digits for the selected unit: Press the
“Confirm” button (configuration parameters, see Chapter “7.3.7
Parameters in the “Unit 1”, “Unit 2”, “Unit 3”, and “Unit 4” Application
Menus”, page 57).
t Select the desired number of display digits and confirm the selection.
t Exit the “Toggle between weight units” function menu.