86 Operating Instructions Cubis
®
MCE
Technical Data
15.7.2 Models MCE524S | MCE524P | MCE324S | MCE324P
MCE524S MCE524P MCE324S MCE324P
Unit Value Value Value Value
Scale interval (d) mg 0.1 0.1 | 0.2 | 0.5 0.1 0.1 | 0.2 | 0.5
Maximum capacity (Max) g 520 120 | 240 | 520 320 80 | 160 | 320
Repeatability at 5% load
Standard deviation of the load values,
tolerance
mg 0.08 0.08 0.08 0.08
Standard deviation of the load values,
typical value
mg 0.04 0.04 0.04 0.04
Repeatability at approx. maximum capacity
Standard deviation of the load values,
tolerance
mg 0.1 0.15 0.1 0.1
Standard deviation of the load values,
typical value
mg 0.05 0.05 0.05 0.05
Linearity deviation
Tolerance mg 0.4 0.5 0.3 0.5
Typical value mg 0.2 0.2 0.2 0.2
Deviation when load is off-center, positions
according to OIML R76
Test weight g 200 200 200 200
Tolerance mg 0.3 0.4 0.3 0.4
Typical value mg 0.2 0.2 0.2 0.2
Sensitivity drift between +10°C and +30°C ppm/K 1 1 1 1
Tare-maximum capacity: Less than 100 %
of maximum capacity
Accuracy class according to Directive
2014/31/EU
I I I I
Verification scale interval (e) according to
Directive 2014/31/EU
mg 1 1 1 1
Minimum load (Min) according to Directive
2014/31/EU
mg 10 10 10 10
Minimum initial weighing according to USP
(United States Pharmacopeia), Chap. 41
Optimum minimum initial weighing mg 82 82 82 82
Typical minimum initial weighing mg 82 82 82 82
Typical stabilization time s 1 1 1 1
Typical measurement time s 3 3 3 3