44 Operating Instructions Cubis
®
MCE
System Settings
7.3 Parameter List
7.3.1 Parameters in the “Setup” Main Menu
Parameters in the “Balance” Submenu
Parameters Setting values Explanation
AMBIENT V.STABLE Sets the ambient conditions to “very stable”: Activates a fast change in the
weight values in the event of a load change with a high output rate.
Recommended for the following work environment:
− Very stable table near the wall
− Closed and calm room
STABLE* Sets the ambient conditions to “stable”.
Recommended for the following work environment:
− Stable table
− Slight movement in the room
− Slight draft
UNSTABL. Sets the ambient conditions to “unstable”: Activates the delayed change in
weight values with a reduced output rate.
Recommended for the following work environment:
− Simple office desk
− Room with moving machinery or personnel
− Slight air movement
V.UNSTBL. Sets the ambient conditions to “very unstable”: Activates a significantly
delayed change in the weight values and long wait for stability with a further
reduction in the output rate.
Recommended for the following work environment:
− Noticeable and slow floor vibrations
− Noticeable building vibrations
− Weighed goods moved
− Very strong air movements
APP FILT. FINAL.RD.* − Activates a filter that enables a fast change in the display for very fast load
changes.
− Display changes with minimal load changes (in the digit range) occur more
slowly.
FILLING Activates a filter that enables a very fast change in the display with minimal
load changes (e.g. when filling containers).
REDUC. Activates a weak but fast filter that always behaves in the same way for load
changes (e.g. when filling automated systems).
OFF Deactivates the active application filter.
STABIL. MAX ACC. Sets the stability to “maximum accuracy”.
V. ACC. Sets the stability to “very accurate”.
ACC.* Sets the stability to “accurate”.
FAST Sets the stability to “fast”.
V. FAST Sets the stability to “very fast”.
MAX.SPEED Sets the stability to “maximum speed”.
* Factory setting