Operating Instructions Cubis
®
MCE 61
Operation
8.7 Adjusting with the isoCAL Function
The device can be automatically internally calibrated and adjusted using the
isoCAL function.
Requirements
− The device is not located in the menu.
− Alphanumeric inputs are not active.
− The load on the weighing pan remains unchanged for 2 minutes.
− The load on the weighing pan amounts to no more than 2% of the
maximum load.
− The device does not register an input for 2 minutes.
If all requirements for starting the isoCAL function are met and one of the
following conditions occurs: The isoCAL function is automatically triggered.
Possible conditions are:
− The ambient temperature has changed since the last
adjustment | calibration.
− The interval time was exceeded (interval time, see Chapter “15.8
Recommended Calibration Weight”, page 90).
− The device has been leveled.
− The device has been disconnected from the power supply since the last
adjustment | calibration (only for conformity-assessed models).
Procedure
t If the isoCAL auto-start function is set (configuration values for the
“ISOCAL” parameter, see Chapter “7.3.1 Parameters in the “Setup” Main
Menu”, page 44):
y If the isoCAL function is triggered according to the aforementioned
conditions: The [isoCAL] button flashes in the operating display.
y In the operating display, a time display counts down from 15 seconds
to 0.
y If no load change or no operation takes place on the device before
the expiration of the time display: The isoCAL function starts.
t If the isoCAL manual start function is set (configuration values for the
“ISOCAL” parameter, see Chapter “7.3.1 Parameters in the “Setup” Main
Menu”, page 44):
y If the isoCAL function is triggered: The [isoCAL] button flashes in the
operating display.
t Press the [isoCAL] button.
y The isoCAL function starts.