90 Operating Instructions Cubis
®
MCE
Technical Data
15.8 Recommended Calibration Weight
MCE225S MCE225P MCE125S MCE125P
Unit Value Value Value Value
External test weight g 200 200 100 100
Recommended accuracy class E2 E2 E2 E2
MCE524S MCE524P MCE324S MCE324P
Unit Value Value Value Value
External test weight g 500 500 300 300
Recommended accuracy class E2 E2 E2 E2
MCE224S MCE124S MCE5203S MCE5203P
Unit Value Value Value Value
External test weight g 200 100 5000 5000
Recommended accuracy class E2 E2 E2 E2
MCE3203S MCE2203S MCE2203P MCE1203S
Unit Value Value Value Value
External test weight g 3000 2000 1000 1000
Recommended accuracy class E2 E2 E2 E2
MCE623S MCE623P MCE323S MCE5202S
Unit Value Value Value Value
External test weight g 500 500 200 5000
Recommended accuracy class E2 E2 E2 E2
15.9 isoCAL Function
15.9.1 Models MCE225S | MCE225P | MCE125S | MCE125P | MCE324S | MCE324P |MCE224S |
MCE124S | MCE2203S | MCE2203P | MCE1203S
Unit Value
isoCAL is triggered by the following criteria:
In the event of a temperature change K 1.5
After a time interval h 12
After successful leveling