262 FLOWSIC100 Flare · Operating Instructions · 8013344/11L2/V 2-5/2018-10 · © SICK Engineering GmbH
Specifications
Subject to change without notice
Fig. 191 Wiring diagram for FLSE100-EX/FLSE100-EXRE
Vervielfaeltigung dieser Unterlage sowie Verwertung und Mitteilung ihres
Inhaltes unzulaessig, soweit nicht ausdruecklich zugestanden. Zuwiderhand-
lungen sind strafbar und verpflichten zu Schadenersatz (LitUrhG, UWG, BGB).
Alle Rechte fuer den Fall der Patenterteilung oder GM-Eintragung vorbehalten.
A4_Formblatt_ME10.CDR
SICK Engineering GmbH
Bergener Ring 27
01458 Ottendorf-Okrilla
Verteildatum:
Page
Norm-Typ/DIN
Ersatz für:
Änderung
gepr.
gez.
Tag
Name
Ind. Tag Name
Maßstab:
Werkstoff
Ersetzt durch:
Gepr.
Ursprung:
2008-08-05
Kochan
MKO
E_41944
4of5
Division1 / Zone 1 Installation
1. In the US install in accordance with the NEC (ANSI/NFPA 70)
2. In Canada install in accordance with the CEC part 1
Class I, Division 1, Groups B, C and D, Temp.Code T6,
Class I, Division 2, Groups A, B, C and D, Temp.Code T6
Class I, Zone 1: Ex/AEx d IIB + H2, Temp.Code T6,
Class I, Zone 2: Ex/AEx nA IIC, Temp.Code T6
MASTER
Ultrasonic Transducer
Zone separating wall e.g. pipeline or duct
FLANGE
GND
+24V
A
B
SLAVE
GND
+24V
A
B
Temperature range
and pressure range
see Marking plate
Conduit size 1/2" NPT
AWG 12
AWG 12
AWG 16
AWG 16
AWG20upto16
AWG20upto16
-50°C < Tamb <55°C, ambient pressure
Manufacturer Drawing FLSE100-EX-6/-RE-6
Power supply
T
Data Interface
erminals +24V, GND
15...28 V dc, 500mA max.
(RS485)
Terminals A, B
+/- 5 V, 500mA max.
Division2 / Zone 2 Installation
1. This equipment is suitable for installation in Class I, Division 2,
Grouos A, B, C and D hazardous locations or nonhazardous
locations only.
2. In the US install in accordance with the NEC (ANSI/NFPA 70)
3. In Canada install in accordance with the CEC part 1
4. WARNING: Explosion Hazard - Do not connect or disconnect
this equipmentn unless power has been switched off or the
area is known to be nonhazardous
5. WARNING: Explosion Hazard - Substitution may impair suitabilty
for Class I Division 2
2009-08-18
TYPE 6, IP 65/67
SINGLE SEAL
Kochan
MKO
A
2009-12-04
2010-08-31
B
MKO
Kochan