Iso-center with ultrasound 6 - 9
Siemens AG SPL1-130.840.02 Page 9 of 16 MODULARIS Uro Plus
Medical Solutions Rev. 03 04.05 CS PS 24
• Make the following basic settings at the ultrasound unit:
Fig. 14 Fig. 15
Value: Adjust using: German English French Spanish
- 6 dB (C/Fig. 14)
Set trackball to SI
(white background),
then actuate and
set value with control
SI Output Sortie Salida
20 dB (B/Fig. 14) Knob B EV Gain Gain Ampl
3 (A/Fig. 14)
Set trackball to FV
(white background),
then actuate and
set value with control
FV Edge Côté Contorno
60 (D/Fig. 14) Knob ------- ------- ------- -------
60 dB (E/Fig. 14)
Set trackball to DR
(white background),
then actuate and
set value with control
DR Dyn Rng DR Cam Din
1
D
B
A
E
C