SDS1000X-E User Manual
28
WWW.SIGLENT.COM
Bandwidth Limit
Set the bandwidth limit to reduce display noise. For example, the signal under test is a
pulse with high frequency oscillation.
When the bandwidth limit set to Full, the high frequency components of the signal
under test can pass the channel.
When the bandwidth limit set to 20M, the high frequency components that exceed 20
MHz are attenuated.
Press the CH1 button on the front panel; then press the BW Limit softkey to select Full or
20M. The default setup is Full. When bandwidth limit is enabled, the character B will be
displayed in the channel label at the right side of the screen.
SDS1000X-E has full BW with all V/div settings including 500uV/div to 2mV/div.
Probe
Set the probe attenuation factor to match the type of the probe that you are using to
ensure correct vertical readouts.
Press the CH1 button on the front panel; then press the Probe softkey and turn the
Universal Knob to select the desired value and push the knob to confirm. The default
setup is 1X.
The current probe attenuation factor is displayed in the channel label at the right side of
the screen. You can also press the Probe softkey continuously to switch the probe
attenuation factor.
The table shows the probe attenuation factor
Menu
Attenuation Factor
0.1X
0.1 : 1
0.2X
0.2 : 1
0.5X
0.5 : 1
1X
1 : 1
2X
2 : 1

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the SIGLENT TECHNOLOGIES SDS1104X-E and is the answer not in the manual?

SIGLENT TECHNOLOGIES SDS1104X-E Specifications

General IconGeneral
BrandSIGLENT TECHNOLOGIES
ModelSDS1104X-E
CategoryTest Equipment
LanguageEnglish

Related product manuals