Operation Spectroquant
®
Pharo 300
104
ba75703d07 04/2014
2 Insert the cell with the reference
sample.
The first single measurement for
the Reference absorbance is car-
ried out.
The following data is displayed as
the result:
– The measured absorbance
from the (last) single measure-
ment.
– The median from all single mea-
surements carried out up to
now.
3 If necessary, carry out further sin-
gle measurements for the forma-
tion of the median with [Next
meas.]
or
discard the last single measure-
ment with [Discard].
4 To accept the median value, press
[Apply].
The photometer is ready to mea-
sure.
The reference absorbance is dis-
played in the top right corner dur-
ing absorbance or transmission
measurement.
Reference absorbance
16.04.07 9:52
To start measurement, insert cell or press
<START/ENTER>
489 nm
10 mm
Reference absorbance
16.04.07 9:52
Last measured absorbance
0.232
Median
0.232 (1 Measurement(s))
489 nm
10 mm
Next meas.
Discard Apply
Absorbance
16.04.07 9:52
Reference::
To start measurement, insert cell or press
<START/ENTER>
489 nm
10 mm
Setup Wavelength Transmission Reference