4.14.3 Printing measurement datasets
This section describes how to print measurement datasets of the measuring
modes, Concentration, Absorbance / % Transmission, and Special / Multi
wavelengths.
By means of sample printouts, the printed information is described below:
Concentration
and Special / Multi
wavelengths mode
Structure of the lines from left to right:
1st line:
[Consecutive no.] [Date] [Time] [Method name] [Measured value] [Unit]
[Citation form] [Dilution] [ID or "AutoStore"]
2nd line (long version only):
[User] [Reagent blank value] [Date of blank value measurement]
[Time of blank value measurement] [Lot ID of blank value measurement]
[AQA1: label] [AQA1: record no.] [AQA2: label] [AQA2: record no]
Note
Optional elements (e.g. dilution or ID) are output only if they were really used
for measurement or storage.
Absorbance / %
Transmission
mode
Structure of the lines from left to right:
1st line:
[Consecutive no.] [Date] [Time] [Wavelength] [Measured value]
["Absorbance" or "Transmission" mode ] [ID or "AutoStore"]
2nd line (long version only):
[User] [Value of reference absorbance] [Date of reference absorbance] [Time
of reference absorbance] [AQA1: label] [AQA1: record no.]
Note
Optional elements (e.g. ID or reference absorbance) are output only if they
were really used for measurement or storage.
21 05.06.07 14:05:41 14541 844 mg/l COD Inlet
Administrator 0.005 02.06.07 11:02:13 2 AQA1: 9 AQA2: 14
14 05.06.07 11:25:01 445 nm 0.609 Absorbance AutoStore
Administrator 0.133 02.06.07 09:59:01 AQA1: 9