3.4.6 12 V-Adapter
With the 12 V-Adapter you can operate the Spectroquant
®
Pharo 300 spec-
trophotometer on the move and independent of the local power supply.
To do so, a 12 V power supply such as a commercial 12 V portable power
source or a 12 V car battery is required.
Safety instructions For operation with an external battery, follow the safety instructions of the
battery.
Make sure the power source is suitable for operation of the spectrophotome-
ter (see technical data of the power source and technical data of the spectro-
photometer).
Operating time with
battery
The maximum operating time depends on various factors:
Battery (e.g. nominal capacity, condition, age)
Operating mode of the spectrophotometer (e.g. frequency of measure-
ments)
Photometer (instrument type)
Example Operating time with a type 12 V / 19 Ah battery in optimum condition: approx.
16h
Note
The spectrophotometer consumes electricity even while it is in standby
mode.
We recommend to disconnect the 12 V-Adapter if you do not use the photom-
eter during battery operation.
Technical data
12 V-Adapter
12 V power supply unit
(e.g. portable power
source or car battery )
Spectroquant
®
Pharo 300
12 V-Adapter
Cable length 2 m
max. voltage 12 V
max. current 8 A