EasyManua.ls Logo

Spectroquant Pharo 300 - Photometer Monitoring (AQA1)

Default Icon
456 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Spectroquant
®
Pharo 300 Operation
163
ba75703d07 04/2014
4.15.2 Photometer monitoring (AQA1)
At least one set of test standards such as Spectroquant
®
PhotoCheck or Cer-
tiPUR
®
is required for the photometer monitoring.
The administrator specifies which test standard has to be used as the mini-
mum requirement for the AQA1 monitoring.
The extent of the monitoring can be enlarged with further test standards.
Note
Settings for AQA checks are only available for users of the user group,
administrator.
Every registered user can carry out the AQA check (see also section 4.16.1).
Spectroquant
®
PhotoCheck
The PhotoCheck consists of 12 test standards in duplicate, 2 zero cells and
2 cells to check the barcode reader. Each PhotoCheck package contains a
lot dependent test certificate with all nominal values (absorbances) and toler-
ances of the test standards. These values are entered in the photometer dur-
ing the configuration of the AQA1 check.
CertiPUR
®
test
standards
Each CertiPUR
®
standard is provided with a lot dependent test certificate
with all nominal values (absorbances) and tolerances of the test standards.
The values were preset in the factory.
Note
Observe the shelf life of the test standards. The values in the photometer
always have to be checked when a new package of test standard is used. If
necessary, adjust the values at the photometer.
Overview of the
photometer
monitoring
Photometer monitoring (AQA1) consists of the following parts:
Configuring settings in the AQA1 setup menu.
Activate AQA1
Specify AQA1 Interval
Activate/deactivate the meter lock for missing or expired AQA1 check
Define the extent of the AQA1 monitoring by activating or deactivating
the individual test standards.
Enter the nominal values, tolerances and lot numbers for the individual
test standards
Carrying out the AQA1 check. The photometer compares the results with
the nominal values while taking into account the tolerances.
The steps are described in detail below.

Table of Contents