EasyManua.ls Logo

Spectroquant Pharo 300 - 4.5 Measuring in Concentration mode

Default Icon
456 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Operation Spectroquant
®
Pharo 300
66
ba75703d07 04/2014
4.5 Measuring in Concentration mode
4.5.1 Measuring cell tests with barcode
Display if the measured value is not within the measuring range (see section
Inserting a cell with barcode starts
a measurement.
1 Open the cell shaft cover.
2 Close the inner turn-up lid.
If the inner turn-up lid is opened
too wide, a message prompts you
to close the inner turn-up lid.
3 Insert the barcoded round cell in
the round cell shaft so it touches
the bottom. When doing so, align
the line mark with the notch at the
front of the round cell shaft.
The photometer selects the
method based on the bar code
and automatically starts measure-
ment.
4 Further options:
Select a different citation form
with [Citation form],
(e.g. NH
4
<–> NH
4
-N).
Select a different measuring
unit with [Unit],
(e.g. mg/l <–> mmol/l).
Make further settings such as
dilution or blank value measure-
ments with [Setup] (see section
4.5.6).
<HOME>
Concentration
Line mark
Barcode

Table of Contents