Pharo 300 Operation
65
ba75703d07 04/2014
Validity of the zero
adjustment
The data of the zero adjustment is stored in the photometer separately for
each cell type. As long as the data is valid, it is automatically used again after
a temporary change to a different cell type. The validity depends on the
respective mode:
* After the wavelength or method respectively was temporarily exited the photometer dis-
plays that a zero adjustment is available and the time it was carried out. You can then
decide whether to use this zero adjustment or carry out a new zero adjustment.
When to repeat the
zero adjustment?
We recommend to repeat the zero adjustment in the following cases:
If the photometer was subject to mechanical stress such as strong shock
or transport
If the ambient temperature changed by more than 5 °C since the last zero
adjustment
At least once per week
If a new cell type (different manufacturer, different glass type is used)
Basically each time you want to measure with the highest possible accu-
racy.
Mode Validity of the zero adjustment
Concentration (permanently
programmed methods)
Till the next zero adjustment
Absorbance / % Transmission Till the next zero adjustment with the
same wavelength *
Concentration (user-defined
methods) and
Special / Multi wavelengths
Till the next zero adjustment for the same
method *
Kinetics Till another kinetic profile is loaded
Till the Kinetics mode is exited or the pho-
tometer is switched off