Spectroquant
®
Pharo 300 Operation
133
ba75703d07 04/2014
Example of a
kinetic recording
(*.csv file)
Line 1 - explanations:
Column Value Explanation
1 6 Version of the file format for the CSV file
2 4001 Profile number
3 1 Measurement of absorbance (0) or transmis-
sion (1)
4 1 Measurement once per interval (0) or as often
as possible (1)
5 525 Wavelength (in nm)
6 1280913092 Start time (internal data format)
7 59 Duration (in sec)
8 5 Interval time (in sec)
9 1 Scaling automatic (0) or manual (1)
10 0.000 Minimum for manual scaling
11 0.301 Maximum for manual scaling
12 0 Enzymatic activity Off (0) or On (1)
13 1.000 Factor for enzymatic activity
14 µkat Unit of enzymatic activity
15 2 Decimal points for enzymatic activity
6|4001|1|1|525|1280913092|59|5|1|0.000|0.301|0|1.000|µkat|2
Device: Serial number:Software: User:
Pharo 300 09130512 1.30-Merck-1.60 Administrator
Start time Wavelength [nm]
04.08.2010 11:11 525
Time [s] Absorbance
0 0,092
5 0,077
10 0,073
15 0,069
.. .....