Spectroquant
®
Pharo 300 Operation
141
ba75703d07 04/2014
For each export procedure you can select the location where the PC-read-
able files (*.csv, *.pdf) should be stored: either to the photometer (Internal
DataB folder) or an external memory (USB memory). On an external mem-
ory, the data are stored in the pHotoLab_6600 directory.
The files stored in the photometer (Internal DataB folder) can later be trans-
ferred to a connected PC or to an external memory (USB memory).
Measurement data Save, back up, export
Concentration,
Absorbance / % Transmis-
sion
Special / Multi wavelengths
Measurement datasets of these measuring
modes are first stored in the measured value
memory of the photometer (1000 memory
locations) with <STORE> or AutoStore.
The measured value memory is available
from the Measurement data memory menu.
Here you can view, filter and export into a PC-
readable file (*.csv) the stored measurement
datasets (<STORE>).
Csv files of these measuring modes cannot
be reimported to the photometer.
Measurement datasets of these measuring
modes can also be stored to a pdf file (see
section 4.11.11).
Spectrum
Kinetics
You can store and export measurement data
of these measuring modes directly as a PC-
readable file (*.csv) with <STORE>.
Csv files of these measuring modes can be
reimported and displayed on the photometer.
Measurement data of these measuring
modes can also be stored to a pdf file (see
section 4.11.11).
AQA records You can store and export measurement data
of these measuring modes directly as a PC-
readable file (*.csv) with <STORE>.
Csv files of records cannot be reimported to
the photometer.
Measurement data of these measuring
modes can also be stored to a pdf file (see
section 4.11.11).
User-defined methods / pro-
files
Method data and profile data are stored and
exported with the Exchange methods/profiles
function in the <HOME>/General setup
menu.