Pipette 1.0 ml of Cd-1
into a test tube.
Add 10 ml of the sample
with pipette and mix.
Add 0.20 ml of Cd-2 with
pipette and mix.
Measuring 0.01 0.500 mg/l Cd 10-mm cell
range: 0.005 0.250 mg/l Cd 20-mm cell
0.0020 0.1000 mg/l Cd 50-mm cell
Expression of results also possible in mmol/l.
Add 1 level green
microspoon of Cd-3
and dissolve the solid
substance.
Reaction time:
2 minutes
Transfer the solution into
a corresponding cell.
Select method with
AutoSelector.
Place the cell into the
cell compartment.
Cadmium
101745
Test
Quality assurance:
To check the measurement system (test reagents,
measurement device, and handling) ready-f
or-use
cadmium standard solution Certipur
®
, Cat.No. 119777,
concentration 1000 mg/l Cd, can be used after diluting
accordingly.
Important:
For the determination of total cadmium a pretreatment
with Crack Set 10C
, Cat.No. 114688 or Crack Set 10,
Cat.No. 114687, and thermoreactor is necessary.
Result can be expressed as sum of cadmium (
Cd).
Check the pH of the
sample, specified range
:
pH 3 – 11.
If required, add dilute
sodium hydroxide
solution or sulfuric acid
drop by drop to adjust
the pH.
Release 06/2014 - Spectroquant
®
UV/VIS Spectrophotometer Pharo 300