Chloride
101807
Test
Release 06/2014 - Spectroquant
®
UV/VIS Spectrophotometer Pharo 300
Quality assurance:
To check the measurement system (test reagents, meas-
urement device, and handling) ready-f
or-use chloride
standard solution Certipur
®
, Cat.No. 119897, concentra-
tion 1000 mg/l Cl
-
, can be used after diluting accordingly.
Pipette 0.20 ml each of
Cl-1 into two test tubes.
Add to one tube 10 ml of
the sample with pipette
and mix.
Reaction time:
10 minutes
Transfer both solutions
into two separate
50-mm-cells.
Select method with
AutoSelector.
Place the blank cell into
the cell compartment.
Add to each tube 0.20 ml
of Cl-2 with pipette and
mix.
Check the pH of the
sample, specified range:
pH 3 – 11.
If required, add dilute
ammonia solution or
nitric acid drop by drop
to adjust the pH.
Measuring 0.10 – 5.00 mg/l Cl 50-mm cell
range: Expression of results also possible in mmol/l.
Add to the second tube
10 ml of of distilled water
(Water for analysis
EMSURE
®
, Cat.No.
116754, is recommend-
ed) with pipette and mix.
(Blank cell)
Place the cell
containing
the sample
into the cell
compartment.
Configure the photometer
for blank-measurement.