Measuring range: 0.5 – 10.0 mg/l Co 10-mm cell
Attention! The measurement is carried out at 495 nm in a 10-mm rectangular cell against a blank, prepared
from distilled water (Water for analysis EMSURE
®
, Cat.No. 116754, is recommended) and the
reagents in an analogous manner.
Cobalt in water
Application
Important:
The exact composition and preparation of the reagents 1,
2, and 3 used are given in the corresponding application,
which also includes further inf
ormation on the method
employed. This application can be downloaded directly at
www.analytical-test-kits.com.
Release 06/2014 - Spectroquant
®
UV/VIS Spectrophotometer Pharo 300
Add 2.0 ml of
reagent 2 with pipette.
Add 1.0 ml of
reagent 3 with pipette,
fill to the mark with dis-
tilled water, and mix
thoroughly.
Pipette 4.0 ml of the
sample into a 25-ml vol-
umetric flask,
fill to the mark with
distilled water and
mix thoroughly.
Add 0.25 ml of
reagent 1 with pipette.
Place the cell into the
cell compartment. Select
method no. 305.
Transfer the solution into
a rectangular cell.