EasyManua.ls Logo

Spectroquant Pharo 300 - Page 300

Default Icon
456 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Filter sample solution
through a membrane
filter with 0.45 µm pore
size.
Notes:
Filtered sample =
true color.
Unfiltered sample =
apparent color.
Transfer the solution
into a corresponding
cell.
Place the cell into the
cell compartment, select
method no. 32.
Color Hazen
(Platinum-Cobalt Standard Method)
analogous to APHA 2120B, DIN EN ISO 6271-2, Water Research Vol. 30, No. 11, 2771-2775, 11996
Release 06/2014 - Spectroquant
®
UV/VIS Spectrophotometer Pharo 300
Quality assurance:
To check the measurement system (measurement device,
handling) ready-for-use Platinum Cobalt Color Ref
erence
Solution (Hazen 500) Certipur
®
, Cat.No. 100246, concentra-
tion 500 mg/l Pt, can be used after diluting accordingly.
Measuring 1 - 500 mg/l Pt/Co 1 - 500 mg/l Pt 1 - 500 Hazen 1 - 500 CU 340 nm 10-mm cell
range: 1 - 250 mg/l Pt/Co 1 - 250 mg/l Pt 1 - 250 Hazen 1 - 250 CU 340 nm 20-mm cell
0.2 - 100.0 mg/l Pt/Co 0.2 - 100.0 mg/l Pt 0.2 - 100.0 Hazen 0.2 - 100.0 CU 340 nm 50-mm cell

Table of Contents