Pipette 5.0 ml of the
sample into a test tube.
Add 2.0 ml of Hy-1 with
pipette and mix.
Measuring 0.02 – 2.00 mg/l N
2
H
4
10-mm cell
range: 0.01 – 1.00 mg/l N
2
H
4
20-mm cell
0.005 – 0.400 mg/l N
2
H
4
50-mm cell
Expression of results also possible in mmol/l.
Hydrazine
109711
Test
Reaction time:
5 minutes
Transfer the solution into
a corresponding cell.
Select method with
AutoSelector.
Place the cell into the
cell compartment.
Important:
To measure in the 50-mm cell, the sample volume and
the volume of the reagents hav
e to be doubled for each.
Alternatively, the semi-microcell, Cat.No. 173502, can be
used.
Quality assurance:
To check the measurement system (test reagents,
measurement device, and handling) a h
ydrazine standard
solution must be prepared from Hydrazinium sulfate GR,
Cat.No. 104603 (see section “Standard solutions”).
Check the pH of the
sample, specified range:
pH 2 – 10.
If required, add dilute
sodium hydroxide
solution or sulfuric acid
drop by drop to adjust
the pH.