Check the pH of the
sample, specified range:
pH 2 – 7.
If required, add dilute
sodium hydroxide
solution or sulfuric acid
drop by drop to adjust
the pH.
Pipette 7.0 ml of the
sample into a reaction
cell
, close with the screw
cap,
and mix.
Add 2 drops of Mn-1K,
close the cell with the
screw cap, and mix.
Measuring 0.10 5.00 mg/l Mn
range: Expression of results also possible in mmol/l.
Manganese
100816
Cell Test
Add 3 drops of Mn-2K,
close the cell with the
screw cap, and mix.
Quality assurance:
To check the measurement system (test reagents,
measurement device, and handling) w
e recommended
to use Spectroquant
®
CombiCheck 30, Cat.No. 114677.
Ready-for-use manganese standard solution Certipur
®
,
Cat.No. 119789, concentration 1000 mg/l Mn, can also
be used after diluting accordingly.
To check for sample-dependent effects the use of addition
solutions (e.g. in CombiCheck 30) is highly recommended.
Reaction time:
2 minutes
Reaction time:
5 minutes
Place the cell into the
cell compartment. Align
the mark on the cell with
that on the photometer.
Release 06/2014 - Spectroquant
®
UV/VIS Spectrophotometer Pharo 300