Molybdenum
119252
Test
Swirl the cell to dissolve
the solid substance.
Add 1 powder pack of
Molybdenum HR2,
close with the screw
cap, and dissolve the
solid substance.
Add 1 powder pack of
Molybdenum HR3 and
close with the screw cap.
Measuring 0.5 – 45.0 mg/l Mo 20-mm cell
range: 0.8 – 75.0 mg/l MoO
4
20-mm cell
1.1 – 96.6 mg/l Na
2
MoO
4
20-mm cell
Expression of results also possible in mmol/l.
Add 1 powder pack of
Molybdenum HR1,
close with the screw
cap, and dissolve the
solid substance.
Pipette 10 ml of the
sample into into a emp-
ty round cell (Empty
cells, Cat.No. 114724).
Reaction time:
5 minutes, measure
immediately.
Quality assurance:
To check the measurement system (test reagents,
measurement device, and handling) a ready-f
or-use
molybdenum standard solution Certipur
®
, Cat.No. 170227,
concentration 1000 mg/l Mo, can be used after diluting
accordingly.
Release 06/2014 - Spectroquant
®
UV/VIS Spectrophotometer Pharo 300
Transfer the solution into
a rectangular cell.
Select method with
AutoSelector.
Place the cell into the
cell compartment.