Nickel in electroplating baths
Inherent color
Pipette 5.0 ml of the
sample into an empty
round cell (Empty cells,
Cat.No. 114724).
Add 5.0 ml of sulfuric
acid 40 %, close the cell
with the screw cap, and
mix.
Place the cell into the
cell compartment.
Select method no. 57.
Measuring 10 120 g/l Ni 10-mm cell
range: 5.0 60.0 g/l Ni 20-mm cell
2.0 24.0 g/l Ni 50-mm cell
Transfer the solution into
a corresponding rectan-
gular cell.
Release 06/2014 - Spectroquant
®
UV/VIS Spectrophotometer Pharo 300