EasyManua.ls Logo

Spectroquant Pharo 300 - Page 379

Default Icon
456 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Pipette 5.0 ml of the
sample into a test tube.
Transfer the solution into
a corresponding cell.
Select method with
AutoSelector.
Place the cell into the
cell compartment
.
114842
Test
Phosphate
Determination of orthophosphate
Measuring 1.0 30.0 mg/l PO
4
-P
3.1 – 92.0 mg/l PO
4
2.3 – 68.7 mg/l P
2
O
5
10-mm cell
range: 0.5 15.0 mg/l PO
4
-P 1.5 46.0 mg/l PO
4
1.1 – 34.4 mg/l P
2
O
5
20-mm cell
Expression of results also possible in mmol/l.
Add 1.2 ml of PO
4
-1 with
piette and mix.
Quality assurance:
To check the measurement system (test reagents,
measurement device, and handling) ready-f
or-use
phosphate standard solution
Certipur
®
, Cat.No. 119898,
concentration 1000 mg/l PO
4
3–
, can be used after
diluting accordingly.
Check the pH of the
sample, specified range:
pH 0 10.
If required, add dilute
sulfuric acid drop by
drop to adjust the pH.
Release 06/2014 - Spectroquant
®
UV/VIS Spectrophotometer Pharo 300
Important:
For the determination of total phosphorus = sum of
orthophosphate,
polyphosphate and organophosphate
either Phosphate Cell Test, Cat. No. 114543, 114729, and
100673 or Phosphate Test, Cat. No. 114848 in conjunction
with Crack Set 10/10C, Cat. No. 114687 resp. 114688 can
be used.

Table of Contents