100885
Cell Test
Measuring 10 300 mg/l Na
range: Expression of results also possible in mmol/l.
Sodium
in nutrient solutions
Add 0.50 ml of the
sample with pipette,
close the cell with the
screw cap, and mix.
Pipette 0.50 ml of Na-1K
into a reaction cell and
mix.
Place the cell into the
cell compartment. Align
the mark on the cell with
that on the photometer.
Reaction time:
1 minute
Quality assurance:
To check the measurement system (test reagents,
measurement device, and handling) ready-f
or-use
chloride standard solution Certipur
®
, Cat.No. 119897,
concentration 1000 mg/l Cl
-
(corresponds to 649 mg/l Na),
can be used after diluting accordingly (see section “Stan-
dard solutions”).