Spectroquant Pharo 300 - Page 402

Default Icon
456 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
101746
Test
Measuring 1.0 – 60.0 mg/l SO
3
10-mm cell
range: 0.8 48.0 mg/l SO
2
10-mm cell
Expression of results also possible in mmol/l.
Shake vigorously
to dissolve the solid
substance.
Add 5.0 ml of distilled
water with pipette and
mix.
Add 2.0 ml of the sample
with pipette and mix.
Place 1 level grey micro-
spoon of SO
3
-1 into a
dry test tube.
Add 3.0 ml of SO
3
-2 with
pipette.
Sulfite
Transfer the solution into
a cell.
Select method with
AutoSelector.
Place the cell into the
cell compartment.
Reaction time:
2 minutes
Quality assurance:
To check the measurement system (test reagents,
measurement device, and handling) a sulfite standard
solution m
ust be prepared from sodium sulfite GR,
Cat.No. 106657 (see section “Standard solutions”).
Check the pH of the
sample, specified range
:
pH 4– 9.
If required, add dilute
sodium hydroxide
solution or sulfuric acid
drop by drop to adjust
the pH.