Check the pH of the
sample, specified range
:
pH 3 9.
If required, add dilute
sodium hydroxide
solution or hydrochloric
acid drop by drop to
adjust the pH.
Pipette 1.0 ml of the
sample into a reaction
cell, close with the screw
cap, and mix.
Reaction time:
3 minutes
100961
Cell Test
Place the cell into the
cell compartment. Align
the mark on the cell with
that on the photometer.
Add 1.0 ml of H-1K with
pipette, close the cell
with the screw cap, and
mix.
Measuring 5 – 215 mg/l Ca
range: 0.7 – 30.1 °d
0.9 – 37.6 °e
1.2 – 53.7 °f
Measuring 7 – 301 mg/l CaO
range: 12 – 537 mg/l CaCO
3
Expression of results also possible in mmol/l
and also in mg/l Mg .
Total Hardness
Determination of total hardness
Release 06/2014 - Spectroquant
®
UV/VIS Spectrophotometer Pharo 300
Quality assurance:
To check the measurement system (test reagents,
measurement device, and handling) a freshly prepared
standard solution
can be used (see section “Standard
solutions”).