EasyManua.ls Logo

Spectroquant Pharo 300 - Page 437

Default Icon
456 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Release 06/2014
Spectroquant
®
photometers NOVA 60 and Pharo photometers
Instructions for the Preparation of Standard Solutions
Standard solution of ozone analogous to
DIN EN ISO 7393
Preparation of a KlO
3
stock solution:
Dissolve 1.006 g of KlO
3
in 250 ml of distilled water in a calibrated or
conformity-checked 1000-ml volumetric flask. Subsequently make up to
the mark with distilled water.
Preparation of a KlO
3
/Kl standard solution:
Transfer 14.80 ml of the KlO
3
stock solution to a calibrated or confor-
mity-checked 1000-ml volumetric flask, add approx. 1 g of Kl and make
up to the mark with distilled water.
1 ml of this solution is equivalent to 0.010 mg of ozone.
Preparation of the ozone standard solution:
Pipette 20.0 ml (full pipette)
KlO
3
/Kl standard solution into a calibrated
or conformity-checked 100-ml volumetric flask, add 2.0 ml of H
2
SO
4
0.5 mol/l, leave to stand for 1 min, and then add NaOH 2 mol/l dropwise
(approx. 1 ml) until the solution just loses its color. Subsequently make
up the solution to the mark with distilled water.
The concentration of the solution is 2.00 mg/l ozone.
Stability:
The KlO
3
stock solution remains stable for 4 weeks when stored in a
cool place (refrigerator). The KlO
3
/Kl standard solution can be used for
5 hours when stored in a cool place (refrigerator). The diluted ozone
standard solution is not stable and must be used immediately.
Reagents required:
1.02404.0100 Potassium
iodate, volum.
standard
1.05043.0250 Potassium iodide
for analysis
EMSURE
®
1.09072.1000 Sulfuric acid
0.5 mol/l
Titripur
®
1.09136.1000 Sodium hy-
droxide solution
2 mol/l
Titripur
®
1.16754.9010 Water for
analysis
EMSURE
®
Standard solution of phenol
Preparation of a standard solution:
Dissolve 1.00 g of phenol GR with distilled water in a calibrated or con-
formity-checked 1000-ml volumetric flask and make up to the mark with
distilled water.
The standard solution pre pared according to this procedure has a con-
centration of 1000 mg/l phenol.
Further investigational concentrations may be prepared from this stan-
dard solution by diluting accordingly with distilled water.
Stability:
When stored in a cool place (refrigerator), the standard solution of
1000 mg/l remains stable for one week. The diluted standard solutions
(inves tigational concentrations) must be used immediately.
Reagents required:
1.00206.0250 Phenol GR for
analysis
1.16754.9010 Water for
analysis
EMSURE
®

Table of Contents