Operation Spectroquant
®
Pharo 300
78
ba75703d07 04/2014
3 Insert the cell with a suitable blank
sample.
The first single measurement for
the sample blank value takes
place.
The following data is displayed as
the result:
The measured absorbance
from the (last) single measure-
ment.
The median from all single mea-
surements carried out up to
now.
4 If necessary, carry out further sin-
gle measurements for the forma-
tion of the median with [Next
meas.]
or
discard the last single measure-
ment with [Discard].
5 To accept the median value, press
[Apply].
The photometer is ready to mea-
sure.
The use of the sample blank value
is indicated by [SB] in the top right
corner of the display.