Parallel Impulse, Fast and Slow detectors for the measurements with A, B, C, Z and LF
frequency filters
Two measurement ranges 25 dBA LEQ ÷ 123 dB Peak (Low) and 30 dBA LEQ ÷ 140 dB Peak
(High).
1/1 Octave real-time analysis (option). Eleven 1/1-octave filters with center frequencies from
16.0 Hz to 16 kHz (meeting Class 1 requirements of IEC 61260-1:2014) available simultaneously
with three user definable profiles for broadband measurements (SLM and SEM), time history
logging and audio recording
1/3 Octave real-time analysis (option). Thirty-five 1/3-octave filters with center frequencies from
8 Hz to 20 kHz (meeting Class 1 requirements of IEC 61260-1:2014) available simultaneously
with three user definable profiles for broadband measurements (SLM and SEM), time history
logging and audio recording
Audio signal recording (option), trigger and continuous mode, 12 kHz and 24 kHz sampling rate,
WAV format
Reverberation time RT60 analysis function (option) for 1/1 octave bands or 1/3 octave bands and
three total RMS levels (A, C and Z weighted) in accordance with the ISO 3382 standard.
STIPA (speech transmission index for public address systems) measurements of electro-acoustic
and acoustic environments effects that affect the speech intelligibility in the room acoustics and/or
public address systems.
1.2 GENERAL FEATURES OF SV 971A
Sound Level Meter in extremely small pocket size body
Noise measurements meeting Class 1 IEC 61672-1:2013 accuracy
Two overlapping wide measurement ranges
1/1 & 1/3 octave real-time frequency analysis (option)
Dosimeter function for personal noise monitoring in the workplace (option)
Audio signal recording (option)
Reverberation time function (option)
Speech transmission index function (option)
Audio records on demand, created before or after measurement, added to a measurement file
Statistical analysis with up to 10 percentile values
Time-history with two logging step intervals
Automated calibration start and save
Free-field & diffuse-field measurements
Integration measurement run time programmable up to 24 h
Wireless connectivity with low energy Long Range Bluetooth
®
Smart (4.2) interface
Setup editor available with Supervisor or SvanPC++ software
Super contrast colour OLED display
Bluetooth
®
for remote control by the BA Assistant smartphone applications
Wide range of temperature operating conditions
Very handy, light weight and robust pocket size case
Easy and friendly user interface for quick start and stop
1.3 ACCESSORIES INCLUDED
ACO 7152 prepolarised ½” condenser microphone with nominal sensitivity 32 mV/Pa
SV 18A microphone preamplifier
SC 158 USB type C to USB type A cable
SA 22 foam windscreen
batteries four AAA type