SV 971A User Manual
62
8 AUXILIARY SETTINGS – Auxiliary Setup
The Auxiliary Setup section provides additional
functions that allow, for instance, customization of the
device interface to a specific user requirement and are
not directly related to the hardware components of the
instrument.
To open the Auxiliary Setup section, press the <Menu>
key, select the Auxiliary Setup position and press
<Enter>.
<ENT>
The Auxiliary Setup section contains following positions:
Language allowing you to select the language of the user interface;
Factory Set allowing you to restore default, factory settings;
Comments allowing you to define the file name for recording voice comments. This position
is available only in the case of the Advanced interface mode;
Leq & Lav allowing you to set the mode of displaying the Leq and Lav results. This position
is available only in the Dosimeter function and in the case of the Advanced
interface mode;
Warnings allowing you to enable/disable warnings to be displayed during the normal
operation of the instrument.
8.1 SELECTING USER INTERFACE LANGUAGE – LANGUAGE
The Language position enables selecting the language
of the user interface.
If after turning the instrument on, an unknown language
interface appears on the display, the user can reset the
instrument with three <Shift/Enter/Start> keys pressed
together during the switching on of the device. After this,
the instrument will go back to the default setup with the
English interface.
<ENT>
8.2 RESTORING FACTORY SETTINGS – FACTORY SETTINGS
The Factory Set. position enables restoring default
settings of the instrument.
Factory settings can be restored also with three
<Shift/Enter/Start> keys pressed together.
<ENT>
After restoring the factory settings, the instrument will
ask you whether to keep the last calibration. If you select
No the factory calibration will be restored and the new
calibration record stating the Factory calibration of the
instrument will be created in the Calibration History list
(see Chapter 3.2.2).
<ENT>