155
Group name
#1
code
#7
code
#S
code
Code description
Measure trigger gradient
O
FK
x - gradient [dB] (1 100); default
10dB/(trigger period)
Auto-Run
AS
#7,AS,<e>,<HH>,<MM>,<hh>,<mm>,<dW>,<m
R>;
where:
<e> - On (e=1), Off (e=0),
<HH> - hour of the measurement start,
<MM> - minutes of the measurement start,
<hh> - hour of the measurement stop,
<mm> - minutes of the measurement stop,
<dW> - day of week in which the measurement
will be done defined as a sum of flags:
b0 - Monday
b1 - Tuesday
b2 - Wednesday
b3 - Thursday
b4 - Friday
b5 - Saturday
b6 - Sunday
<mR> - maximum number of the measurement
days,
MR
<e> - On (e=1), Off (e=0)
MJ
<HH> - hour of the measurement start
MK
<MM> - minutes of the measurement start
ML
<hh> - hour of the measurement stop
MM
<mm> - minutes of the measurement stop
MN
<dW> - day of week
MO
<mR> - maximum number of the measurement
days
Table A.3 Calibration and microphone settings
Group name
#1
code
#7
code
#S
code
Code description
Calibration factor
Q
AJ
nn.nn - calibration factor [dB] represented as real
number (-10.00 10.00)
Calibration history
CH
#7,CH;
Get number of records calibration history.
Response format:
#7,CH,n;
n number of records calibration history,
#7,CH,n;
Get n record from calibration history.
Response format:
#7,CH,n,cT,hh,mm,ss,DD,MM,YYYY,cF,c
L;
n record number in the history of
calibration,

Table of Contents