4
REMOTE INTERFACES
Device Specific Commands
4-18 TP04300 Series Interface & Applications Manual
CLER Clear device-specific (reported by EROR?) errors.
NOTE: After sending this command, wait 4 seconds before sending
another command.
COOL Turn the compressor on or off.
COOL 1 – turn the compressor on
COOL 0 – turn the compressor off
NOTE: There is a 60 second delay between the time that the
compressor is turned on and the System is ready to operate.
COOL? Read COOL on/off state.
CYCC Set the cycle count.
CYCC nnnn -- where nnnn is the number (1 - 9999) of cycles to do.
CYCC? Read the number of cycles to do.
CYCL Start/stop cycling.
CYCL 1 – start
CYCL 0 – stop
NOTE: When all cycles have been completed or when cycling was
stopped on failure, it is necessary to send a CYCL 0 command to
reset the system.
CYCL? Read the cycle number (current value if cycling, last value if not).
NOTE: If Version 1 compatibility mode is enabled, CYCL? returns
the number of fully completed cycles.
CYCO Turn the display of the cycling feature parameters on the operator
screen on or off.
CYCO 1 – display cycling parameters
CYCO 0 – don’t display cycling parameters
CYCO? Read CYCO on/off state.
CYCP Pause or restart cycling.
CYCP 1 – pause cycling
CYCP 0 – restart cycling
CYCP? Read CYCP pause/run state
DLOG DLOG “filename” turns on datalogging to the file “filename.”
DLOG 0 turns datalogging off.
NOTE: The file name should be in quotes.
NOTE: If the file already exists, it will automatically be
overwritten.
DLOG? Return the name of the file if datalogging is active, or “None” if
datalogging is off.
DSNS Set the DUT sensor type.
DSNS n -- where n is 0-4
0 – no DUT sensor
1 – type T thermocouple
2 – type K thermocouple
3 – RTD
4 – diode
Command Description