4
REMOTE INTERFACES
Device Specific Commands
4-24 TP04300 Series Interface & Applications Manual
SRST Reset (force) the System to the Cycle screen without raising the test
head.
NOTE: Any device-specific errors are reset. The upper and lower
temperature limits and certain other values are reset. Setpoint
number 5 becomes the active setpoint.
NOTE: After sending this command, wait 4 seconds before sending
another command.
STIM set the shutdown timer to shutdown the system after n minutes
where n is 0 - 12960.
NOTE: Setting a value of “0” will disable the shutdown timer.
STIM? queries the time (in minutes) before the timer will shutdown the
system.
STND Raise or lower the test head (same as HEAD).
STND 1 -- put head down
STND 0 -- put head up
NOTE: Sending this command when the head is locked will NOT
cause an error, but the head will not actually move.
TECR? Read the temperature event condition register.
bit 7 – datalogging on
bit 6 -- not used
bit 5 -- stopped cycling ("stop on fail" signal was received)
bit 4 -- end of all cycles
bit 3 -- end of one cycle
bit 2 -- end of test (test time has elapsed)
bit 1 -- not at temperature
bit 0 -- at temperature (soak time has elapsed)
TEMP? Read the main temperature, in 0.1 °C increments.
NOTE: This query returns air temperature when in air-control
mode, or DUT temperature when in DUT-control mode. Use
TMPA? to always return air temperature, and TMPD? to always
return DUT temperature, regardless of mode.
TESE Set the temperature event status enable (mask) register.
TESE nnn -- where nnn is 0 – 255
NOTE: See TESR? for the meaning of each bit in the mask.
TESE? Read the temperature event status enable (mask) register.
Command Description