EasyManua.ls Logo

Thermo Scientific Applied Biosystems QuantStudio - Page 57

Thermo Scientific Applied Biosystems QuantStudio
126 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Observation Possible Cause Action
The chip was not completely
covered with Immersion
Fluid immediately after
loading (while still on the
chip loader or heated block)
Note: (QuantStudio
3D
Instrument firmware version
1.1 or earlier) Yellow around
the edges is a strong
indication of evaporation.
Add Immersion Fluid as soon as the
loading blade is off of the chip.
Ensure that the entire chip is covered
with Immersion Fluid, even the edges.
Note: When adding Immersion Fluid,
the uid can overow into the chip
base, but do not allow it to spread
onto the sides of the chip lid.
Immersion Fluid was not
applied to the chip
immediately after loading
(evaporation of the PCR
reaction).
Debris or other
contaminants are present on
the chip lid during imaging.
No action required.
The AnalysisSuite
Software can
compensate for small quantities of
debris or other contaminants on the
chip.
Chapter 6 Troubleshooting
Troubleshooting chip images using the Chip View
6
QuantStudio
3D Digital PCR System User Guide
57

Table of Contents

Related product manuals