Observation Possible Cause Action
Condensation may have
been present on the chip
during imaging.
IMPORTANT! Condensation may
not be visible on the chip.
•
If condensation is not visible – If
possible, read the chip again
after it has warmed to room
temperature. Once equilibrated,
the data quality, ag, and
separation of amplied versus
non-amplied peaks should
improve noticeably.
•
If condensation is visible – Using
a low-lint wipe, remove any
condensation or Immersion Fluid
from the surface of the chip by
wiping in one direction. If
needed, use a low-lint wipe
sprayed with isopropanol to
remove any dried residue.
The loading blade was not
seated correctly within the
loader arm.
Before pressing the loading button,
conrm that the loading blade is rmly
seated and ush against the loader
arm.
Chapter 6 Troubleshooting
Troubleshooting chip images using the Chip View
6
58
QuantStudio
™
3D Digital PCR System User Guide