User's Guide _______________________________________________________________________
78 ___________________________________________________________________ M211107EN-C
Checksum CSUM has to be calculated using CRC16 method (see section
CRC16 Checksum on page 222).
For example, command
FSA
&DO STATUS
036B
prints the status messages in frames (equal to message 3), whereas
command
FSA
&DO MEAS_SYNC
44DA
resets the measurement sequence. Command
FSA
&DO NAME
28D4
displays the device type, device name given by the user, and ID.
Command
FSA
&DO SYSTEM
6EEE
lists system information, such as type, ID, software version, hardware
modules, and serial numbers.