12-3
IM 01E30D01-01EN
<12. OUTLINE>
Totalizer Display Function (*2):
Theowrateiscountedonecountatatimeaccord-
ing to the setting of totalization pulse weights. For
forwardandreverseowmeasurementfunctions,
thetotalizedvaluesoftheowdirection(forwardor
reverse)andtheowdirectionaredisplayedonthe
indicator together with the units. The difference of
totalized values between the forward and reverse
owratecanbedisplayed.Totalizationforthere-
verseowrateiscarriedoutonlywhen“Forwardand
reverseowmeasurementfunctions”isselected.
Damping Time Constant (*2):
The time constant can be set from 1.0 second to
200.0 seconds (63% response). The default is 5
seconds.
Whenthedampingtimeisshort,theoutputuctu-
ates. Set the time to 5 seconds or more for control
loop.
Span Setting Function (*2):
Spanowscanbesetinunitssuchasvolumeow
rate,massowrate,time,orowratevalue.The
velocity unit can also be set.
Volume Flow Rate Unit: kcf, cf, mcf, Mgal (US), kgal
(US), gal (US), mgal (US), kbbl (US)*,
bbl(US)*,mbbl(US)*,μbbl(US)*,Ml
(megaliter), m
3
, kl (kiloliter), l (liter), cm
3
Mass Flow Rate Unit (Density must be set): klb
(US), lb (US), t (ton), kg, g
Velocity Unit: ft, m (meter)
Time Unit: s (sec), min, h (hour), d (day)
* “US Oil” or “US Beer” can be selected.
Pulse Output (*1)(*2):
Scaled pulse is output one by one according to the
setting of pulse weight.
PulseWidth:Duty50%orxedpulsewidth(0.05,
0.1, 0.5, 1, 20, 33, 50, 100, 200, 330,
500, 1000, 2000 ms) can be selected.
Output Rate: 0.0001 to 10,000 pps (pulses per
second)
Multi-range Functions (*1)(*2):
Automatic range switching
Whentheowrateexceeds100%oftherange,
transition to the next range (up to two ranges) is
carried out automatically. Range switching can be
conrmedbythestatusoutputandontheindicator.
Forward and Reverse Flow Measurement Func-
tions (*1)(*2):
Flows in both forward and reverse directions can be
measured.Thereverseowmeasurementcanbe
conrmedbythestatusoutputandontheindicator.
Totalization Switch (*1)(*2):
The status output is carried out when a totalized val-
ue becomes equal to or greater than the set value.
Preset Totalization (*1)(*2):
The parameter setting enables a totalized value to be
preset to a setting value or zero.
Alarm Selection Function (*2):
AlarmsareclassiedintotheSystemAlarms(hard-
ware failure), Process Alarms (such as Signal
OverowandAdhesionAlarm),SettingAlarms,and
Warnings. Process Alarms and Setting Alarms can be
activated or deactivated for each item. The current
output for an alarm can be selected from the follow-
ing settings. If any System Alarm occurs, turn the
power off and back on again to return to the normal
condition.
System Alarm: 21.6 mA or more, 3.2 mA or less
Process Alarm, Setting Alarm:
21.6 mA or more, 20.5 mA, HOLD
(xedtothecurrentvaluebefore
the alarm), 4 mA, 3.8 mA, 3.2
mA or less.
The default settings of each alarm are as follows:
Standard
Optional
code C1
System Alarm
21.6 mA or more 3.2 mA or lessProcess Alarm
Setting Alarm
Note: In the case of style: S1 with optional code C1, the
current output is set up to 3.8 mA during Process
Alarms and Setting Alarms.
NE-107 Alarm Message (*2):
Alarmsareclassiedinto4categoriesbyNAMUR
NE-107 and can be displayed.
-F: Failure
-C: Function check
-S:Outofspecication
-M: Maintenance required
Alarm Output (*1)(*2):
Alarms are generated only for the items selected
via the ‘Alarm Selection Function’ if relevant failures
occur.
Self Diagnosis Functions (*2):
When an alarm is output, details of the System
Alarms, Process Alarms, Setting Alarms, and Warn-
ingsaredisplayedtogetherwiththespecicdescrip-
tion of corresponding countermeasures.
Resultsofmeanowmeasurementduringagiven
period, etc. can be checked by using parameters.
FlowRateUpper/LowerLimitAlarms(*1)(*2):
Ifaowratebecomesoutofthepredetermined
range, the alarm output is generated.
Electrode Adhesion Diagnosis Function (*1) (*2):
This function enables checking of the adhesion level
of insulating substances to the electrodes. Depend-
ingonthestatusofadhesion,usersarenotiedbya
warning or an alarm via status outputs.
While adhesion diagnosis is being carried out (ap-
proximately 5 minutes), a current of 4 mA is output
becausetheowmeasurementisnotperformed.
Adhesion diagnosis should only be carried out when
theuidvelocityiscompletelyzerobyclosingthe
valve.
When adhesion diagnosis is carried out, change the
controllooptothemanualmoderst.
Data Security during Power Failure:
Data (parameters, totalization value, etc.) storage by
EEPROM. No back-up battery required.
Low Cut (*2):
In this function, the values of the current output along
with LCD indication, totalization, and pulse, which are
correspondingtosettingspanof0to20%,arexed
at0%(includingreverseow).Thedefaultsettingis
3%.
When the Low Cut point is small, the incorrect output
mayoccurattheowrateofzero.SettheLowCut
point to 3% or more. If the span is small, the damping
timeisshortortheuidislowconductivity,theincor-
rectoutputmayeasilyoccurattheowrateofzero.