<6. PARAMETER DESCRIPTION>
6-42
IM 01E30D01-01EN
(1) Display and output condition for system alarms
Alarm description
Alarm
output
4-20mA
output
Totalization Pulse Display unit
Alarm
record
Normal Closed
(ON)
Normal Normal Normal Display Mode No
10 μPFault Microprocessor
(CPU) failure
Open
(Off)
Fixed at
3.2mA or less
otherwise
21.6mA or
more (*)
Indetermination Indetermination Indetermination Indetermination
11 EEPROM
Fault
EEPROM failure Open
(Off)
Fixed at
3.2mA or less
otherwise
21.6mA or
more (*)
Stopped Stopped Alarm Mode
(display of
system Alarm
message)
Recorded
12 Sub uP Fault Substitute
mcroprocessor
(CPU) failure
13 EXPwrFault Extitation power
circuit failure
14 A/D(S) Fault A/D converter failure
(Signal)
15 A/D(I) Fault A/D converter failure
(Excitation current)
16 Analog Fault Analog circuit failure
17 Coil Open Flowtube coil is
open-circuit
18 Coil Short Flowtube coil is
short-circuit
19 Excite Error Excitation circuit
failure
20 Pulse Error Discrepancy of
pulse count
21 EEPROM
Dt
EEPROM default
values
Note: • Alarmisoutputonlywhen[F10:DO Function] is set to “Alarm Output.”
• Theoperationwhen“Open(Off)Act”issetfor[F11:DO Active Mode] is shown in above table.
*: The output value is performed in accordance with the setting of the burnout switch. For information about this switch, read Subsection
11.3.1.
(2) Display and output condition for process alarms
Alarm description
Selection
(parameter
number)
Alarm
output
4-20 mA
output
Totalization
Pulse
output
Displayunit
Alarm
record
30 SigOverow Input signal error YES
(G26)
Open
(Off)
Fixed Stopped Stopped Alarm Mode
(Message)
Recorded
NO
(G26)
Closed
(On)
Continuous
(*)
Continuous
(*)
Continuous
(*)
Display
Mode
No
31 Empty Pipe Flowtubeisnotlled
withuid
YES
(G27)
Open
(Off)
Fixed Stopped Stopped Alarm Mode
(Message)
Recorded
NO
(G27)
Closed
(On)
Continuous
(*)
Continuous
(*)
Continuous
(*)
Display
Mode
No
32 H/L Alm H/L Alarm YES
(G28)
Open
(Off)
Normal
operation
Normal
operation
Normal
operation
Alarm Mode
(Message)
No
NO
(G28)
Closed
(On)
Display
Mode
33 Adhesion Alm Electrode adhesion
alarm
YES
(G29)
Open
(Off)
Fixed Stopped Stopped Alarm Mode
(Message)
Recorded
NO
(G29)
Closed
(On)
Continuous
(*)
Continuous
(*)
Continuous
(*)
Display
Mode
No
Note: • larmisoutputonlywhenF10: DO Function is set to “Alarm Output.”
• The operation when “Open (Off) Act” is set for F11: DO Active Mode is shown in above table.
• 4-20mAoutputupontheoccurrenceofanalarmwillbexedatthevalueselectedwithG25: 4-20mA Process Alarm.
*: Although outputs are continuous, output values are not guaranteed.