6243/44 DC Voltage Current Source/Monitor Operation Manual
9.2.1 Voltage/Current Source
9-13
Overall accuracy: Including calibration accuracy, 1-day stability, temperature coefficients, and linearity
1-day stability: When the AC power and load are constant:
Voltage source /
Voltage limiter
Range Overall accuracy (% of setting + V)
320 mV
3.2 V
20 V
0.03 + 300 V
0.03 + 600 V
0.03 + 6 mA
Current source /
Current limiter
Range Overall accuracy (% of setting + A + A Vo / 1 V )
320 A
3.2 mA
32 mA
320 mA
3.2 A
10 A
0.03 + 100 nA + 3 nA
0.03 + 1 A + 30 nA
0.03 + 10 A + 300 nA
0.05 + 100 A + 3 A
0.06 + 1 mA + 30 A
0.1 + 10 mA + 300 A
Vo: Compliance voltage (0 V to 20 V)
Voltage source /
Voltage limiter
Range Overall accuracy (% of setting + V)
320 mV
3.2 V
20 V
0.01 + 150 V
0.01 + 300 V
0.01 + 3 mV
Voltage source /
Voltage limiter
Range Overall accuracy (% of setting + A + A Vo / 1 V )
320 A
3.2 mA
32 mA
320 mA
3.2 A
10 A
0.015 + 42 nA + 2 nA
0.01 + 420 nA + 20 nA
0.01 + 4.2 A + 200 nA
0.015 + 42 A + 2 A
0.03 + 420 A + 20 A
0.08 + 4.2 mA + 200 A
Vo: Compliance voltage (0 V to 20 V)