CALCulate[1|2|3|4]:LIMit:UPPer:CLEar[:IMMediate] command
Deletes the upper limit line from the specified display.
Command Syntax: CALCulate[1|2|3|4]:LIMit:UPPer:CLEar[:IMMediate]
Example Statements: OUTPUT 711;"Calc3:Limit:Upp:Clear:Imm"
OUTPUT 711;"CALC:LIMIT:UPP:CLEAR"
Attribute Summary: Option: not applicable
Synchronization Required: no
Preset State: not applicable
SCPI Compliance: instrument-specific
Description:
To delete an upper limit line, send CALC:LIM:UPP:CLE. To delete a lower limit line, send
CALC:LIM:LOW:CLE.
You can delete part of a limit line if it consists of segments. See the CALC:LIM:UPP:SEGM:CLE
command for information about deleting a segment of the upper limit line.
CALCulate
6-44