CALCulate[1|2|3|4]:LIMit:UPPer:MOVE:Y command
Moves all segments of the upper limit line up or down in the specified trace box.
Command Syntax: CALCulate[1|2|3|4]:LIMit:UPPer:MOVE:Y <number>|<bound>
<number> ::= a real number (NRf data)
limits: -9.9e37:9.9e37
(depend upon current Y-axis value)
<bound> ::= MAX|MIN
Example Statements: OUTPUT 711;":calc:lim:upper:move:y -3.86779e+37"
OUTPUT 711;"Calc:Lim:Upper:Move:Y -6.89317e+37"
Attribute Summary: Option: not applicable
Synchronization Required: no
Preset State: not applicable
SCPI Compliance: instrument-specific
Description:
This command increments or decrements all segments in a upper limit by the specified value along the
Y-axis. The value is unitless and assumes the current vertical/division unit (returned with
DISP:TRAC:Y:PDIV? UNIT).
To specify trace box A, send CALC1:LIM:UPP:MOVE:Y. To specify trace box B, send
CALC2:LIM:UPP:MOVE:Y. To specify trace box C, send CALC3:LIM:UPP:MOVE:Y. To specify
trace box D, send CALC4:LIM:UPP:MOVE:Y. If a trace specifier is not included in the command, the
trace specifier defaults to trace box A.