CALCulate[1|2|3|4]:MARKer:DTABle:X:LABel command/query
Loads a label for the selected entry in the data table.
Command Syntax: CALCulate[1|2|3|4]:MARKer:DTABle:X:LABel ‘<STRING>’
<STRING> ::= ASCII characters - 0 through 255
maximum number of characters: 16
Example Statements: OUTPUT 711;"calc1:mark:dtab:x:lab ‘Fundamental’"
OUTPUT 711;"calc3:mark:dtable:x:label ‘ Carrier’"
Query Syntax: CALCulate[1|2|3|4]:MARKer:DTABle:X:LABel?
Return Format: “<STRING>”
Attribute Summary: Option: not applicable
Synchronization Required: no
Preset State: not applicable
SCPI Compliance: instrument-specific
Description:
To select the X value entry in the data table, use the CALC:MARKER:DTAB:X:SELECT[:POINT]
command. The entry must exist in the data table before you assign a label to it.
CALCulate[1|2|3|4] specifies the data table. Send CALC1 to specify the data table appearing in trace box
A, CALC2 to specify B, CALC3 to specify C, and CALC4 to specify D. If you do not send the optional
specifier, the command defaults to the data table appearing in trace box A.
To display the data table, use the DISPlay[:WINDow[1|2|3|4]]:DTABle[:STATe] command.
CALCulate
6-64

Table of Contents

Other manuals for Agilent Technologies 35670A

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies 35670A and is the answer not in the manual?

Agilent Technologies 35670A Specifications

General IconGeneral
BrandAgilent Technologies
Model35670A
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals