The total number of fixed poles or zeros in the table cannot be greater than the number entered under the curve
fit setup key.
The value entered for number of poles cannot be less than the total fixed poles in the table.
The value entered for number of zeros cannot be less than the total fixed zeros in the table.
This data register contains data with non-uniform X-axis spacing and therefore cannot be edited.
This feature cannot be used when the marker is at zero.
This key is only valid when XDCR units are ON and the Y-axis of the active trace is an amplitude.
This marker function is not valid for arbitrarily spaced data.
This marker function is only valid for complex data.
This marker function is only valid for frequency domain data.
This marker function is only valid for frequency or order domain data.
This marker function is only valid for frequency, order, or octave domain data.
This marker function is only valid for frequency-response data.
This marker function is only valid for octave domain data.
This marker function is only valid for power or linear spectrum Trace Data and MAGNITUDE, REAL, or
IMAGINARY Trace Coordinates.
This marker function is only valid for time domain data.
This menu is only valid for single channel frequency or time data when XDCR unit is ON and is not set to
USER.
To use a 102.4 kHz span the antialias filter must be turned off for channel 1.
To use the TRACE RPM feature, data must be taken in RPM STEP ARM mode.
Too many rpm steps for order tracking.
Trace contains invalid data.
Trace has too many points to convert to a limit line, only the first 1024 points will be used.
UP ONLY auto range not available in Swept Sine instrument mode.
Unit Conversion not possible due to incompatible RPM PROFILE.
Use CALC:MARK:POS command to move marker on ORBIT Data and NYQUIST or POLAR Trace Coord.
Value entered cannot be zero.
Value entered must have same sign as current marker value.
WATERFALL STEPS must be > 1.
WATERFALL cannot be turned ON with FRNT/BACK display formats.
Waterfall markers require WATERFALL ON.
Waterfall operation not available with measurement running.
When the 102.4 kHz span is being used the antialias filter for channel 1 cannot be turned on.
X-axis units cannot be changed for this data.
Y unit selection is meaningless for the current combination of Meas Data and Trace Coord selections.
Y-axis scale matching not possible with current Trace Coord selections.
dB REF SETUP requires Trace Coordinate to be dB MAGNITUDE and XDCR UNIT to be Pascals, USER, or
OFF.
dbSPL choice requires Trace Coordinate to be dB MAGNITUDE and XDCR UNIT to be Pascals.
dbm choice requires Trace Coordinate to be dB MAGNITUDE and XDCR UNIT to be OFF or V.
-211: Trigger ignored.
Bus trigger ignored when trigger type is not GPIB trigger.
Trigger received when not waiting for trigger.
Error Messages
Execution Errors
C-6

Table of Contents

Other manuals for Agilent Technologies 35670A

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies 35670A and is the answer not in the manual?

Agilent Technologies 35670A Specifications

General IconGeneral
BrandAgilent Technologies
Model35670A
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals